Fetch:array:marker:power, Fetch:array:temperature:average, Fetch:array:temperature:current – Boonton 4540 Peak Power Meter User Manual User Manual

Page 159: Fetch:cw:power, Fetch:array:marker:power? -25, Fetch:array:temperature:average? -25, Fetch:array:temperature:current? -25, Fetch:cw:power? -25, Boonton 4540 series rf power meter, Remote operation

Advertising
background image

Boonton 4540 Series RF Power Meter

Remote Operation

5-25

FETCh:ARRay:MARKer:POWer?

Description:

Returns an array of the current marker measurements for the specified channel. The array consists
of the average, maximum, and minimum power and peak-to-average ratio between the two
markers, powers at both markers, and the ratio of the two markers. Note the peak-to-average ratio
and marker ratio are returned in dB for log units, and percent for linear units.

Syntax:

FETCh[1|2]:ARRay:MARKer:POWer?

Returns:

CC1, Pavg, CC2, Pmax, CC3, Pmin, CC4, PkToAvgRatio, CC5, Pwr@Marker1, CC6,
Pwr@Marker2, CC7, Mrk1/Mrk2 ratio

Where the CCn’s are the measurement condition codes for each measurement..

Valid Modes:

Pulse and Modulated Modes

FETCh:ARRay:TEMPerature:AVERage?

Description:

Returns the averaged sensor temperature for each channel. Note: sensor temperature is only
supported for peak power sensors.

Syntax:

FETCh:ARRay:TEMPerature:AVERage?

Returns:

CC1, Sensor1 Average Temp in degrees C, CC2, Sensor2 Average Temp in degrees C

Where the CCn’s are the measurement condition codes for each measurement..

Valid Modes:

All

FETCh:ARRay:TEMPerature:CURRent?

Description:

Returns the instantaneous sensor temperature for each channel. Note: sensor temperature is only
supported for peak power sensors.

Syntax:

FETCh:ARRay:TEMPerature:CURRent?

Returns:

CC1, Sensor1 Temp in degrees C, CC2, Sensor2 Temp in degrees C

Where the CCn’s are the measurement condition codes for each measurement..

Valid Modes:

All

FETCh:CW:POWer?

Description:

Return current average amplitude reading in channel units.

Syntax:

FETCh[1|2]:CW:POWer?

Returns:

CC, average power (watts, dBm) or average voltage (volts, dBv)

Where CC is the measurement condition code.

Valid Modes:

Modulated and Statistical Modes.