Boonton 4540 Series RF Power Meter

Remote Operation

5-41

OUTPut:INTernal:SIGNal

Description:

Set or return the on/off state of the internal 50 MHz output signal.

Syntax:

OUTPut:INTernal:SIGNal <Boolean>

Argument:

<Boolean> = { 0, 1, OFF, ON }

Valid Modes:

Any

OUTPut:MIO:MODe

Description:

Set/return Multi IO mode. Below is a description of each mode:

Mode

Description

Off:

Disables multi IO output.

Recorder:

Analog output driven by 4540 input signal.

Voltage:

DC voltage output set by user.

Status:

TTL output on events such as an alarm, measurement ready, or
calibration complete.

Trigger:

TTL pulse out on trigger event.

Syntax:

OUTPut:MIO:MODe <character data>

Argument:

<character data> = OFF, STATus, RECOrder, TRIGger, VOLTage

Valid Modes:

Any

OUTPut:[MIO:]RECOroder:FORCe

Description:

Command sets the Multi IO to voltage mode and sets the output voltage to the argument. Query
returns the output voltage if Multi IO mode is in voltage mode or causes error -221 ―Settings
conflict‖ if the Multi IO is a different mode. This command is provided for 4530 legacy
compatibility.

Syntax:

OUTPut:[MIO:]RECOroder:FORCe <numeric_value>

Argument:

<numeric_value> = -10.0 to + 10.0 V

Valid Modes:

Any

OUTPut:[MIO:]RECOroder:MAX

Description:

Set or return the Multi IO recorder output maximum, or full scale (+10.0V) power reference level.
For voltage probes, this is the equivalent power at the current impedance setting.

Syntax:

OUTPut:RECOrder:MAX <numeric_value>

Argument:

<numeric_value> = -200.00 to +200.00 dBm

Valid Modes:

Any