Boonton 4540 series rf power meter, Operation – Boonton 4540 Peak Power Meter User Manual User Manual

Page 118

Advertising
background image

Boonton 4540 Series RF Power Meter

Operation

4-70

4.5.46 System > Calibrator Menu (External 1 GHz - optional)

The Calibrator Menu is used to seclect and control the external 1 GHz RF Calibrator (Model 2530). It can also select the 50
MHz internal calibrator. The external calibrator may be pulse modulated using either a built-in pulse generator, or via a rear-
panel BNC pulse input. Note that this menu does not contain any items related to sensor calibration—it is only for controlling
the calibrator for use as a signal source. For sensor calibration information, refer to Section 3.5 and Section 4.4.2 of this
manual.

Main Pg2 > System > Calibrator > Cal Output

Description:

Enable or disable the RF output of the 1 GHz external
calibrator and select the mode of operation.


Menu Type:

Multiple Choice.

Choices:

Off: Disabled

CW: Provides 1 GHz CW output

Int Pulse: Provides pulse modulation from internal

pulse generator.

Ext Pulse: Provides pulse modulation from external

signal to BNC input connector.

SCPI Command:

OUTPut:EXTernal:SIGNal <Boolean>

Main Pg 2 > System > Calibrator > Set Level

Description:

Set the output of the 1 GHz external calibrator in dBm.


Menu Type:

Numeric value.

Range:

-60.0 dBm to +20.0 dbm (Internal 50 MHz)

-50.0 dBm to +20.0 dbm (External 1 GHz)

SCPI Command:

OUTPut:EXTernal:LEVel:[POWer]
<numeric_value>



Main Pg 2 > System > Calibrator > Sel Calibrator

Description:

Select the active calibrator for calibration and control. ―Int‖ selects the internal 50 MHz
calibrator. ―Ext‖ selects an external 1 GHz calibrator. "Ext" cannot be selected if an
external calibrator is not connected. An error message pop-up will appear.


Menu Type:

Toggle.

Choices:

Int, Ext


SCPI Command:

None.







Fig.4-58 Calibrator Menu

External 1 GHz