4.5.48 System > Calibrator > Pulse - Variable Menu (Ext 1 GHz - opt.)


The pulse menu is specific to the 1 GHz calibrator when it is connected. Control of the internal pulse generator is availble in
2 forms. The Variable mode provides a wide choice of periods and pulse widths, but has some restrictions. Preset mode
provides a limited choice of periods and duty-cycles, but is easy to use.

Main Pg 2 > System > Calibrator > Pulse > Polarity

Description:

Selects the pulse polarity.


Menu Type:

Toggle


Choices:

+, -


SCPI Command:

None.



Main Pg 2 > System > Calibrator > Pulse > Pulse Width

Description:

Sets the pulse width in seconds.


Menu Type:

Numerical Values


Range:

7 usec to 65.535 msec for period ≤ 65.542 msec

and (width + 7 usec) ≤ period. Minimum width

increases with increasing period above 65.544 msec.

SCPI command:

See Pulse Period below.



Main Pg 2 > System > Calibrator > Pulse > Pulse Period

Description:

Sets the pulse width in seconds.


Menu Type:

Numerical Values


Range:

28 usec to 131.077 msec. Period must be an even number of microseconds.


SCPI Command:

OUTPut:EXTernal:PULSe:PERWID <numeric-value> <numeric-value>

< Pulse Period > < Pulse Width >



Main Pg 2 > System > Calibrator > Pulse > Pulse Control


Description:

Selects the pulse control method.


Menu Type:

Toggle.


Choices:

Preset, Variable


SCPI Command:

OUTPut:EXTernal:PUISe:CTRL <character data>