Fetch:temperature:average, Fetch:temperature:current, Fetch:temperature:internal – Boonton 4540 Peak Power Meter User Manual User Manual

Page 164: Fetch:temperature:average? -30, Fetch:temperature:current? -30, Fetch:temperature:internal? -30, Boonton 4540 series rf power meter, Remote operation

Advertising
background image

Boonton 4540 Series RF Power Meter

Remote Operation

5-30

FETCh:TEMPerature:AVERage?

Description:

Return the averaged internal temperature of the specified peak power sensor. Note: If a non-
temperature compensated peak sensor is in use, the 4540 displays a warning message if the
temperature has drifted more than 4C from the autocal temperature.

Syntax:

FETCh[1|2]:TEMPerature:AVERage?

Returns:

CC, Sensor[1|2] Average Temp in degrees C

Where CC is the measurement condition code.

Valid Modes:

Peak Sensors only

FETCh:TEMPerature:CURRent?

Description:

Return the current internal temperature of the specified peak power sensor. Note: If a non-
temperature compensated peak sensor is in use, the 4540 displays a warning message if the
temperature has drifted more than 4C from the autocal temperature.

Syntax:

FETCh[1|2]:TEMPerature:CURRent?

Returns:

CC, Sensor[1|2] Temp in degrees C

Where CCn is the measurement condition code.

Valid Modes:

Peak Sensors only

FETCh:TEMPerature:INTernal?

Description:

Return the current internal temperature of the instrument in degrees Celsius.

Syntax:

FETCh:TEMPerature:INTernal?

Returns:

CC, Instrument Temp in degrees C

Where CCn is the measurement condition code.

Valid Modes:

all

Advertising