Boonton 4540 Series RF Power Meter

Operation

4-15

4.5.9 Channel > Extensions Pg 1 Menu (Modulated Mode)

The extensions menu provides access to more advanced channel options such as trace filtering and channel units.
Additionally access to menu page two is provided. This menu is for sensors in Modulated Mode.


Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Units

Description:

Set the channel units. For power sensors, voltage is
calculated with reference to the sensor input impedance.
Note that for ratiometric results, logarithmic units will
always return dBr (dB relative) while linear units return
percent.

Menu Type:

Pop-up Selection.


Choices:

dBm, Watts, Volts, dBV, dBmV, dBµV


SCPI Command:

CALCulate[1|2]:UNIT <character data>


Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections

Description:

Selects the channel corrections menu.

Menu Type:

Menu Select.


Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Filter State

Description:

Set the current setting of the integration filter on the
selected channel. ―Off‖ provides no filtering, and can
be used at high signal levels when absolute minimum
settling time is required. ―On‖ allows a user-specified
integration time, from 2 milliseconds to 15 seconds.
―Auto‖ uses a variable amount of filtering, which is set automatically by the power meter
based on the current signal level to a value that gives a good compromise between
measurement noise and settling time at most levels.

Menu Type:

Multiple Choice.


Choices:

Off, On, Auto


SCPI Command:

SENSe[1|2]:FILTer:STATe <character data>





Fig. 4-9 Extensions Menu

Pg1 (Modulated Mode)