Output:mio:status:setting, Output:mio:status:source, Output:mio:status:setting -43 – Boonton 4540 Peak Power Meter User Manual User Manual

Page 177: Output:mio:status:source -43, Boonton 4540 series rf power meter, Remote operation

Advertising
background image

Boonton 4540 Series RF Power Meter

Remote Operation

5-43

OUTPut:MIO:STATus:SETTing

Description:

Set/return Multi IO status out setting. Below is a description of each argument:

Argument

Description

ALMACT:

Sets the Multi IO output to +5 VDC when an alarm is currently active.
Multi IO output will be 0 VDC if no alarms are active.

ALMLATCH:

Sets the Multi IO output to +5 VDC when an alarm limit has been
exceeded. Multi IO output will be 0 VDC if no alarm limits have been
exceeded.

MEASRDY:

Multi IO output is 5 VDC when a measurement is ready to be read.

CAL:

Multi IO output is 5 VDC when a calibration is complete.

Syntax:

OUTPut:MIO:STATus:SETTing <character data>

Argument:

<character data> = { ALMACT, ALMLATCH, MEASRDY, CAL }

Valid Modes:

Any

OUTPut:MIO:STATus:SOURce

Description:

Set/return Multi IO source for status out (alarm). Single channel units (Model 4541) only accept
―CH1‖.

Syntax:

OUTPut:MIO:STATus:SOURce <character data>

Argument:

<character data> = { CH1, CH2, BOTH }

Valid Modes:

Any