SENSe:ARRay:CALTemp?

Description:

Returns both channels’ sensor calibration temperatures.

Syntax:

SENSe:ARRay:CALTemp?

Returns:

CC1, Sensor1 cal temp in degrees C, CC2, Sensor2 cal temp in degrees C

Where CCn is the measurement condition code.

Valid Modes:

All

SENSe:CALTemp?

Description:

Returns the selected channel’s sensor calibration temperature.

Syntax:

SENSe[1|2]:CALTemp?

Returns:

CC, Sensor[1|2] cal temp in degrees C

Where CCn is the measurement condition code.

Valid Modes:

All

SENSe:CORRection:CALFactor

Description:

Set or return the frequency calfactor currently in use on the selected channel. Note setting a
calfactor with this command will override the ―automatic‖ frequency calfactor that was calculated
and applied when the operating frequency was set, and setting the operating frequency will
override this calfactor setting.

Syntax:

SENSe[1|2]:CORRection:CALFactor <numeric_value>

Argument:

<numeric_value> = –3.00 to 3.00 dB

Valid Modes:

All

SENSe:CORRection:DCYCle

Description:

Set or return the duty cycle correction factor currently in use on the selected channel.

Syntax:

SENSe[1|2]:CORRection:DCYCle

Argument:

<numeric_value> = 0.01 to 100.00 percent

Valid Modes:

CW Sensor (Modulated Mode)