4.5.28 Cursors Menu (Statistical Mode)

The statistical cursors menu controls the mode and the vertical or horizontal position
of the CCDF graphical measurement cursor. This cursor can operate with either the value
of the normalized CCDF (percent probability) or the power setting as the indepenedent
variable. Whichever axis is chosen as the independent variable, the orthogonal one will
become the calculated result of the CCDF automatically.

Main Pg1 > Stat Mode > Cursors > Cursor Mode

Description:

Select the independent variable of the measurement

cursor. If Percent is selected, power at the cursor's

intersection with the CCDF curve will be measured.

If Power Ref is selected, probability at the cursor's

intersection with the CCDF curve will be measured.

Menu Type:

Toggle


Choices:

Percent, Power Ref


SCPI Command:

None


Main Pg1 > Stat Mode > Cursors > Cursor Pct

Description:

Set the measurement cursor to the desired probability.

This menu box is dimmed if Cursor Mode is set to

Power Ref.


Menu Type:

Numerical Value


Range:

0.000 to 100.000 %


SCPI Command:

MARKer:POSItion:PERcent



Main Pg1 > Stat Mode > Cursors > Cursor Pow Ref

Description:

Set the measurement cursor to the desired power. This

menu box is dimmed if Cursor Mode is set to Percent.

Menu Type:

Numerical Value


Range:

0.000 to 200.000 dBr


SCPI Command:

MARKer:POSItion:POWer


See the following related SCPI commands:

READ:MARKer:CURsor:PERcent?
READ:MARKer:CURsor:POWer?


Fig. 4-39 Cursors Menu