6 initiate and abort commands, Abort, Initiate:continuous – Boonton 4540 Peak Power Meter User Manual User Manual

Page 165: Initiate[:immediate[:all, 6 initiate and abort commands -31, Abort -31, Initiate:continuous -31, Initiate[:immediate[:all]] -31, Boonton 4540 series rf power meter, Remote operation

Advertising
background image

Boonton 4540 Series RF Power Meter

Remote Operation

5-31

5.6.6 INITiate and ABORt Commands

The purpose of the INITiate group of commands is to start and control the process of data acquisition once a measurement
has been configured. Depending on settings, the 4540 RF Power Meter may be commanded to begin either a single
measurement (INITiate:CONTinuous OFF) which stops when complete, or enter a ―free-run‖ mode where data acquisition
occurs continuously (INITiate:CONTinuous ON). The ABORt command terminates any operation in progress and prepares
the instrument for an INITiate command. In some operating modes, the INITiate commands do not actually start
measurements, but rather arm a hardware trigger, which is then used to gate the actual measurements cycle.


ABORt

Description:

Terminates any measurement in progress and resets the state of the trigger system. Note that
ABORt will leave the measurement in a stopped condition with all current measurements cleared,
and forces INITiate:CONTinuous to OFF.

Syntax:

ABORt

Argument:

None

Valid Modes:

Any

INITiate:CONTinuous

Description:

Set or return the data acquisition mode for single or free-run measurements. If
INITiate:CONTinuous is set to ON, the 4540 immediately begins taking measurements
(Modulated, CW and Statistical Modes), or arms its trigger and takes a measurement each time a
trigger occurs (Pulse Mode). If set to OFF, the measurement will begin (or be armed) as soon as
the INITiate command is issued, and will stop once the measurement criteria (averaging, filtering
or sample count) has been satisfied. Note that INITiate:IMMediate and READ commands are
invalid when INITiate:CONTinuous is set to ON; however, by convention this situation does not
result in a SCPI error.

Syntax:

INITiate:CONTinuous <Boolean>

Argument:

<Boolean> = { 0, 1, OFF, ON }

Valid Modes:

Any

INITiate[:IMMediate[:ALL]]

Description:

Starts a single measurement cycle when INITiate:CONTinuous is set to OFF. In Modulated
Mode, the measurement will complete once the power has been integrated for the full FILTer
time. In Pulse Mode, enough trace sweeps must be triggered to satisfy the AVERaging setting. In
Statistical Mode, acquisition stops once the terminal condition(s) are met. In each case, no reading
will be returned until the measurement is complete. This command is not valid when
INITiate:CONTinuous is ON, however, by convention this situation does not result in a SCPI
error.

Syntax:

INITiate[:IMMediate[:ALL]]

Argument:

None

Valid Modes:

Any

Restrictions:

INITiate:CONTinuous must be OFF

Advertising