Display:trace:vcenter, Display:trace:vscale, Display:trace:vcenter -22 – Boonton 4540 Peak Power Meter User Manual User Manual

Page 156: Display:trace:vscale -22, Boonton 4540 series rf power meter, Remote operation

Advertising
background image

Boonton 4540 Series RF Power Meter

Remote Operation

5-22

DISPlay:TRACe:VCENTer

Description:

Set or return the power or voltage level of the horizontal centerline of the graph for the specified
channel in channel units. If a change in the vertical scale causes the center maximum value to be
exceeded, the center will be forced to the maximum value for the new range.

Syntax:

DISPlay:TRACe[1|2]:VCENTer <numeric_value>

Argument:

<numeric_value> = -200.00 to +200.00 dBm for dBm units

10,000 times the vertical scale power/div for watts units

10,000 times the vertical scale volts/div for volts units

Valid Modes:

All

DISPlay:TRACe:VSCALe

Description:

Set or return the power or voltage vertical sensitivity of the trace display in channel units.

Syntax:

DISPlay:TRACe[1|2]:VSCALe <numeric_value>

Argument:

Units = dBm, dBV, dBmV, dBuV , <numeric_value> = range in dB/division

Units = watts

<numeric_value> = range in watts/division

Units = volts

<numeric_value> = range in volts/division

Valid Modes:

All