Boonton 4540 Series RF Power Meter

Remote Operation

5-28

FETCh:MARKer:CURsor:PERcent?

Description:

Return the CCDF cursor y-axis position in percent with respect to the value set by
MARKer:POSItion:POWer (CCDF cursor mode is Power Ref). If CCDF cursor mode is Percent,
returns user setting. See MARKer:POSItion:POWer and MARKer:POSItion:PERcent. See also
Section 6.4 Statistical Mode Automatic Measurements for more information.

Syntax:

FETCh[1|2]:MARKer:CURsor:PERcent?

Returns:

CC, percent CCDF

Where CC is the measurement condition code.

Valid Modes:

Statistical.

FETCh:MARKer:CURsor:POWer?

Description:

Return the CCDF cursor x-axis position in relative power with respect to the value set by
MARKer:POSItion:PERcent (CCDF cursor mode is Percent). If CCDF cursor mode is Power Ref,
returns user setting. See MARKer:POSItion:POWer and MARKer:POSItion:PERcent. See also
Section 6.4 Statistical Mode Automatic Measurements for more information.

Syntax:

FETCh[1|2]:MARKer:CURsor:POWer?

Returns:

CC, relative power (dBr) CCDF

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated.

FETCh:MARKer:DELTa?

Description:

For the specified channel return the difference between MK1 and MK2. The units will be the same
as marker units.

Syntax:

FETCh[1|2]:MARKer:DELTa?

Returns:

CC, (MK1 – MK2)

Where CC is the measurement condition code.

Valid Modes:

All

FETCh:MARKer:MAXimum?

Description:

For the specified channel and marker, return the maximum power or voltage at the marker. The
units are the same as the specified channel.

Syntax:

FETCh[1|2]:MARKer[1|2]:MAXimum?

Returns:

CC, maximum power or voltage at marker

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated.

Advertising