Output:mio:trigout:source, Output:mio:voltage, Output:mio:trigout:source -44 – Boonton 4540 Peak Power Meter User Manual User Manual

Page 178: Output:mio:voltage -44, Boonton 4540 series rf power meter, Remote operation

Advertising
background image

Boonton 4540 Series RF Power Meter

Remote Operation

5-44

OUTPut:MIO:TRIGout:SOURce

Description:

Set/return Multi IO source for trigger out mode. Below is a description of each argument:

Argument

Description

INT1, INT2:

Multi IO output is active when selected internal trigger is active.

EXT:

Multi IO output is active when external trigger is active.

SWEEP:

Pulses when actual trigger event occurs.

Syntax:

OUTPut:MIO:TRIGout:SOURce <character data>

Argument:

<character data> = { INT1, INT2, EXT, SWEEP}

Valid Modes:

Any



OUTPut:MIO:VOLTage

Description:

Set/return Multi IO user set voltage out.

Syntax:

OUTPut:MIO:VOLTage:SOURce <numeric_value>

Argument:

< numeric_value > = -10.0 to + 10.0 V

Valid Modes:

Any