Boonton 4540 Series RF Power Meter

Remote Operation

5-50

READ:MARKer:CURsor:PERcent?

Description:

Return the CCDF cursor y-axis position in percent with respect to the value set by
MARKer:POSItion:POWer (CCDF cursor mode is set to Power Ref). If CCDF cursor mode is
Percent, returns user setting. See MARKer:POSItion:POWer and MARKer:POSItion:PERcent.
See also Section 6.4 Statistical Mode Automatic Measurements for more information.

Syntax:

READ[1|2]:MARKer:CURsor:PERcent?

Returns:

CC, percent CCDF

Where CC is the measurement condition code.

Valid Modes:

Statistical.

READ:MARKer:CURsor:POWer?

Description:

Return the CCDF cursor x-axis position in relative power with respect to the value set by
MARKer:POSItion:PERcent (CCDF cursor mode is set to Percent). If CCDF cursor mode is
Power Ref, returns user setting. See MARKer:POSItion:POWer and MARKer:POSItion:PERcent.
See also Section 6.4 Statistical Mode Automatic Measurements for more information.

Syntax:

READ[1|2]:MARKer:CURsor:POWer?

Returns:

CC, relative power (dBr) CCDF

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated.

READ:MARKer:RATio?

Description:

For the specified channel return the ratio of MK1 to MK2. The units will be dB for logarithmic
units or percent for linear units.

Syntax:

READ[1|2]:Marker:RATio?

Returns:

CC, MK1/MK2

Where CC is the measurement condition code.

Valid Modes:

All