27 term options menu (statistical mode), 27 term options menu (statistical mode) -48, Boonton 4540 series rf power meter – Boonton 4540 Peak Power Meter User Manual User Manual

Page 96: Operation

Advertising
background image

Boonton 4540 Series RF Power Meter

Operation

4-48

4.5.27 Term Options Menu (Statistical Mode)

The statistical mode menu provides for controlling the horizontal axis parameters of the CCDF display, the length of time or
maximum sample count of the CCDF and the action to be taken when the terminal condition is met.

Main Pg1 > Stat Mode > Term Options > Term Action

Description:

Select the action to take when either the terminal

count is reached or the terminal time has elapsed.

Stop - Stop accumulating samples and hold the result.


Restart - Clear the CCDF and begin a new one.


Decimate - Divide all sample bins by 2 and continue.

Menu Type:

Multiple choice


Choices:

Stop, Restart, Decimate


SCPI Command:

TRIGger:CDF:DECimate <Boolean>

Note: For remote operation, start/stop action is

controlled by the INITiate system.



Main Pg1 > Stat Mode > Term Options > Term Count

Description:

Set the terminal sample count for the CCDF.


Menu Type:

Numeric Value


Range:

2 to 4,096 Megasamples


SCPI Command:

TRIGger:CDF:COUNt <Numeric_value>



Main Pg1 > Stat Mode > Term Options > Term Time

Description:

Set the terminal running time for the CCDF.


Menu Type:

Numeric Value


Range:

1 to 3600 seconds


SCPI Command:

TRIGger:CDF:TIMe <Numeric_value>

Fig. 4-38 Term Options

Menu (Statistical)




























Advertising