2 power requirements, 3 connections, 2 power requirements -2 – Boonton 4540 Peak Power Meter User Manual User Manual

Page 26: 3 connections -2, Boonton 4540 series rf power meter, Installation

Advertising
background image

Boonton 4540 Series RF Power Meter

Installation

2-2

Table 2-1 4540 Series Packing List

INSTRUMENT (See also Table 1-1)

4540 Series RF Power Meter

Line Cord
Fuse Kit
Boonton Instruction Manual CD

SENSOR(S) (packaged separately)
Sensor(s)
Sensor Cable(s)
Type N to SMA Adapter (if required)
BOONTON Sensor Manual CD

For bench-top use, choose a clear, uncluttered area. Ensure that there is at least 2" of clearance at the fan air intake on the
rear panel and the exhaust vents on the side panels. Pull-down feet are located on the bottom of the instrument. Rack
mounting instructions are provided with the optional rack mount kit.

2.2 Power Requirements


The 4540 Series is equipped with a switching power supply that provides automatic operation from a 90 to 260 volt, 47 to 63
Hz, single-phase, AC power source. Maximum power consumption is 50W and 70VA. For metric fuse sizes, use the metric
fuse kit supplied. Connect the power cord supplied with the instrument to the power receptacle on the rear panel. See Figure
3-2.

Caution

Before connecting the instrument to the power source, make certain that a 1.0-ampere time delay fuse (type
T) is installed in the fuse holder on the rear panel.

Before removing the instrument cover for any reason, position the input module power switch to off (0 =
OFF; 1 = ON) and disconnect the power cord.



2.3 Connections


Sensor(s)

Connect the sensor that covers the frequency range of the measurement to the CHANNEL 1 sensor
connector on the front (Standard) or rear (Optional) panel, as follows. Connect the sensor to the sensor
cable by aligning the red mark on each part and pressing the connectors together firmly. Connect the sensor
cable to the CHANNEL 1 Input, holding the red mark on the cable connector up. For two-channel
measurements, use the same procedures to connect the second sensor to the CHANNEL 2 Input.


Note

If the sensor connector is not a type N, install the appropriate adapter

(from the accessories kit) on the

calibrator output connector.

Advertising