Setting the Output Form of the Tolerance Comparator

The output form of the tolerance comparator can be set. The following three types are
available.

Setting the output form of the tolerance comparator

1

Hold down the

key, and select "

oPt on

" with the

key.

Set the common settings.

2

Press the

key, and select function

"

d

" with the

key.

Set the output form of the tolerance comparator.

3

Press the

key, and select

"

d-1

" with the

key.

"Hold" is set in this example.

4

Press the

key to register the setting. Press the

key to return to the measurement state.

Function No.

Output form

Function

d-0

nor al

Normal

Outputs according to the tolerance comparator.

d-1

kold

HOLD

Holds the output that is turned ON. Releases the holding when
the measurement value is reset.

d-2

delay

Off-delay

Off-delay of 60ms is applied to the Normal output. Releases
the holding when the measurement value is reset.