Response delay time, Outside dimensions, Response delay time -17 – KEYENCE LK-G Series User Manual

Page 171: Outside dimensions -17

Advertising
background image

6-17

6

LK-G-M-NO6-E

6 Specifications of the LK-G Series

Response Delay Time

■

Measurement response time

This is the period of time until the unit recognizes the change in the target and outputs the
signal. This is calculated by the following formula.

• Measurement response time = measurement average time + refresh rate + internal

processing time (T6 x 4)

■

Power ON reset time

This is the period of time until measurement starts after turning on the power.

• Power ON reset time = approx. 5 sec.

Outside Dimensions

Unit (mm)

■

Controller

LK-G3001V/LK-G3001PV

Refer to "Timing Details" (page 4-16) for the measurement average time, refresh rate (T5),
and sampling rate (T6).

4-M4 screw, depth:6

6

62

117

(65)

33

70

43

9.5

124