Environment settings parameters, Output, Outputting measurement values in external – KEYENCE LK-G Series User Manual

Page 152: Synchronization -22, Environment settings parameters -22, Output -22

Advertising
background image

5-22

5

LK-G-M-NO5-E

5 RS-232C

Outputting Measurement Values in Exter-
nal Synchronization

This section describes how to synchronize the measurement value with the mea-
surement performed by the commands from the input terminal (TIMING input) and
output it from the RS-232C interface.
This method does not use any incoming command, so it is used when the measure-
ment value is only imported or when the unit is connected to a device such as a data
logger that cannot send a command.

Commands for measurement value output and for setting change are available even when
the measurement value output is in use in external synchronization (when the external trig-
ger input is other than OFF).

Do not fix the next measurement value while the current measurement value is outputted
via RS-232C. That next measurement value will not be outputted.

Environment Settings Parameters

Set the environment settings as follows.

Output

The measurement value that is fixed in the external synchronization (TIMING1 and 2 input) is
outputted.

The output content becomes as follows depending on the setting of auto transmission.

Refer to "Setting the Communication Specifications of the RS-232C" (page 3-49) for
Environment settings.

Item

Setting value

Remark

Baud rate

9600/19200/38400/57600/115200 bps

Set the parameter according to the
external devices to be connected.

Parity check

None/Even/Odd

Auto transmission

OUT1/OUT2/OUT1 + OUT2

Select one of them according to
the output method.

Auto transmission

Output contents

OFF

Does not output the measurement values in
external synchronization.

OUT1

Outputs only OUT1.

OUT2

Outputs only OUT2.

OUT1 + OUT2

Outputs OUT1 and OUT2 at the same time.

Reference

Note