Setting the scaling for measurement (calibration), Setting the scaling for measurement, Calibration) -20 – KEYENCE LK-G Series User Manual

Page 78

Advertising
background image

3-20

3

LK-G-M-NO3-E

3 Function Settings

Setting the Scaling for Measurement (Calibration)

The displayed value for the measurement value can be calibrated as desired.

Set the values to display for the desired two points to calibrate.

Head A and head B can be calibrated for each OUT1 and OUT2.

Setting procedure of scaling

This section describes the setting procedure of the following calibration as an example.

1

Hold down the

key, and select "

oUt-1

" with the

key.

OUT 1 is set in this example.

Select " oUt-2 " when setting OUT2.

2

Press the

key, and select function "

b

" with the

key.

Set the scaling.

3

Press the

key, and select "

b-0

".

"Head A" is set in this example.

Select " b-1 " when setting head B.

Setting content

Setting value

Display

Point 1

Original input value 1 -1.0200

A1. nPt

Calibrated displayed
value 1

-1.0000

A1.d SP

Point 2

Original input value 2 0.9900

A2. nPt

Calibrated displayed
value 2

1.0000

A2.d SP

Display

Input

0.99

0.99

-1.02

-1.02

0.99

1.0000

-1.0000

-1.02

Display

Input

Before calibration

After calibration

OUT1

HI

GO

LO

HI

GO

LO

TIM

OUT2

TIM

OUT1

HI

GO

LO

HI

GO

LO

TIM

OUT2

TIM

OUT1

HI

GO

LO

HI

GO

LO

TIM

OUT2

TIM