5-23

5

LK-G-M-NO5-E

5 RS-232C

Timing Chart

When the measurement mode is other than sample hold (trigger 2)

• When the measurement mode is set to normal, the measurement value of when the timing

input turns on is outputted.

• When the measurement mode is peak hold, bottom hold, peak-to-peak hold, average

hold or sample hold (trigger 1), the measurement value that is fixed by turning on the
timing input is outputted.

When the measurement hold is sample hold (trigger 2)

• After the timing input turns on, the measurement value fixed after the sampling period

+t ms is outputted.

The t-time is identified as follows.

Sampling rate

20 µs

t=T2+4 ms

Other than 20 µs t=T2+3 ms

Refer to page 4-16 for T2 (input response time).

TIMING input

Measurement

value

t ms

RS-232C data output

ON
OFF

TIMING input

Measurement

value

Sampling period + t ms

RS-232C data output

ON
OFF