3-27

3

LK-G-M-NO3-E

3 Function Settings

Timing Chart

Normal

Measurement is performed consecutively, and the measurement value is displayed and outputted as
required.

• The ON state of each output indicates the state where the NPN or PNP open-collector output is turned on.
• The ON state of each input indicates the state where it is short-circuited between the terminal and the COM

terminal for NPN type, and the voltage is applied between the terminal and the COM terminal.

• The RS-232C output is synchronized with the timing input when "Auto transmission" in the Environment set-

ting is set other than OFF. (page 3-49)

Refer to "Functions of the Input and Output Signals" (page 4-7) for details of binar y output.

If the RESET input is set to ON when the TIMING input is ON, the comparator standby state (the dis-

play shows "

------

") is displayed until the TIMING input becomes OFF.

• The TIMING input and RESET input can be controlled using the RS-232C (page 5-8).
• There is no difference between the functions of Trigger 1 and Trigger 2.

Measurement value in Normal mode

Internal measurement value

ON

OFF

HI output

t

Comparator standby

(------ display)

HI set value

Measurement value

LO set value

ON

OFF

GO output

ON

OFF

LO output

ON

OFF

TIMING output

ON

OFF

RESET input

RS-232C output

Measurement

value

RESET process

Reference