Setting the Sampling Rate of Measurement Value

This section describes how to set the measurement sampling sycle. If you want the sampling
to be performed more quickly, modify it to perform faster or when the received light is extremely
dim, modify it to perform more slowly.

Setting procedure of the sampling rate

1

Hold down the

key, and select "

oPt on

" with the

key.

Set the common settings.

2

Press the

key, and select function "

A

" by the

key.

Set the sampling cycle.

3

Press the

key, and select

"

A-4

" with the

key.

The sampling rate is set as "500" µs in this example.

4

Press the

key to register the setting. Press the

key to return to the measurement state.

• The sampling rate will be doubled when two heads are connected and the sampling

rate is set as 20 µs.

• Set the sampling rate to other than 20 µs (other than 20 µs and 50 µs when transparent

object 2 is selected in the LK-G85/G80/G87/G82/G155/G150/G157/G152/G405/G400/
G407/G402, and other than 20 µs, 50 µs and 100 µs when transparent object 2 is
selected in the LK-G505/G500/G507/G502) when one of the items from transparent
object, transparent object 2, and multi-reflective object is selected in the
measurement mode of the head (

page 3-7).

• The sampling rate will be up to four times slower when the transparent object 2 is

selected.

• Do not use the measurement value display or received light waveform display of the

setup support software "LK-Navigator" when setting the sampling rate to 20 µs.

Function No.

Function (sampling rate)

A-0

20

20 µs

A-1

50

50 µs

A-2

100

100 µs

A-3

200

200 µs

A-4

500

500 µs

A-5

1000

1000 µs

OUT1

HI

GO

LO

HI

GO

LO

TIM

OUT2

TIM

OUT1

HI

GO

LO

HI

GO

LO

TIM

OUT2

TIM

OUT1

HI

GO

LO

HI

GO

LO

TIM

OUT2

TIM

Note