Measuring with Offset

You can add or subtract any value to or from the displayed value. When the offset value is set, it
can be displayed when the auto-zero is performed.

The offset value is set for the measurement value that underwent a measurement mode process
and the auto-zero process.

Adjusting the master workpiece by using the auto-zero function (auto offset function)
The size of the master workpiece can be set as the master value (reference value), if you
set the size of the master workpiece as the offset value, and then use the auto-zero func-
tion while measuring the master workpiece.

Setting procedure of the offset value

The setting procedure is described below for the case where offset is set for OUT1 as an
example.

1

Hold down the

key, and select "

oUt-1

" with the

key.

OUT1 is chosen for this example.

2

Press the

key, and select function "

F

" with the

key.

Set the offset.

3

Press the

key, and set the offset value with the

and

keys.

"0.03" is set in this example.

4

Press the

key to register the setting. Press the

key to return to the measurement state.

Refer to "List of Default Values and Setting Ranges" (3-15 page) for the setting range and
its restriction.