Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 119

Advertising
background image

5 • Service

Doc. 065291-01 3/09

107

3. On the Chromeleon Wellness Panel under Pump Flow Rate

Calibration, select Reset Cal, or touch

Reset Cal

on the touch

screen.

4. Select

1 mL/min

to set the flow rate and begin pumping deionized

water.

5. Allow the pump to stabilize for 20 minutes.

6. If you are using the Wellness Panel, click

5.00 min

to start the timer.

If you are using the touch screen, start an external timer. Immediately
start collecting water into the tared beaker.

7. Collect the water for exactly 5.00 minutes.

8. Enter the weight of the water (in grams).

9. If you are using the Wellness Panel, click

Calibrate

to download the

value to the ICS-2100.

10. To recheck the value, wait at least 15 minutes, and then repeat the

calibration.

11. If you are using the Wellness panel, click

Log

to record the new

calibration value in the Audit Trail.

12. If you performed the calibration from the ICS-2100 touch screen,

record the new calibration values in the Chromeleon Wellness
database. To do this, open the Wellness Panel (see

Section 5.1.1

) and

click Upload.

Advertising