5 • service, 179 filling the egc co, Mixer with carbonate eluent – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 191

Advertising
background image

5 • Service

Doc. 065291-01 3/09

179

Filling the EGC CO

3

Mixer with Carbonate Eluent

Figure 5-70

illustrates the liquid flow path through the eluent generator

components during initial filling of the EGC CO

3

Mixer with carbonate

eluent.

Figure 5-70. Plumbing for Initial Filling of the EGC CO

3

Mixer with Carbonate Eluent

EGC

DEGAS

EGC

IN

Connect to

Gas Separator

Assembly

TO DEGAS

ELUENT OUT

TO SRS/AES

REGEN OUT

WASTE,

GAS

SEPARATOR

Connect to

Suppressor

Regen Out

Pre-plumbed

Pump Pulse

Damper

to

EGC
OUT

TO DEGAS

ELUENT IN

Back-

pressure

Coil

TO INJ

VALVE IN - P/

ELUENT OUT

To waste

EGC CO

Mixer

3

TO EGC CO3

MIXER IN

TO EGC CO3

MIXER OUT

14 MPa (2000 psi)
at 1.0 mL/min

TO PUMP/

DAMPER

To the EluGen

cartridge inlet port

From the EluGen
cartridge outlet port

Advertising