A • specifications – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 211

Advertising
background image

A • Specifications

Doc. 065291-01 3/09

199

EluGen
Cartridges

Eluent Concentration Range

Comment

K

2

CO

3

/EPM

Electrolytic pH
Modifier

0.1 to 15 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 2.0 mL/min flow
X = 15/flow

The total of the eluent
concentrations from the
EluGen cartridge and the
EPM must not exceed
the specified range. The
EPM concentration must
not exceed 10 mM.

KOH/KOH
KOH/MSA
KOH/NaOH
MSA/MSA
MSA/NaOH
NaOH/NaOH

0.1 to 50 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
X = 50/flow

The eluent concentration
range for each cartridge
is cut by 50%.

LiOH/LiOH

0.1 to 40 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
X = 40/flow

The eluent concentration
range for each cartridge
is cut by 50%.

Table A-7. Eluent Concentration Ranges for Linked

Dual-Cartridge Configurations

Advertising