Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 196

Advertising
background image

ICS-2100 Ion Chromatography System

184

Doc. 065291-01 3/09

5. Verify that the suppressor and the auxiliary electrolytic device (if

installed) are off.

6. Verify that the

Automatically turn on with pump

option is not

selected on the touch screen

SUPPRESSOR

page (see

Section B.9

).

Disabling this option lets you turn on the pump, without
automatically turning on the suppressor.

7. Set the pump flow rate to the rate required for the application.

8. Turn on the pump.

9. Verify that the system backpressure is between 14 and 16 MPa (2000

and 2300 psi). If necessary, adjust the pressure by adding or removing
backpressure tubing.

10. Set the

EGC-1

and

EGC-2

concentrations to the values required for

your application. For example, for a 3.5 mM K

2

CO

3

/1.0 mM KHCO

3

mixture, set the

EGC-1

concentration to 3.5 mM and the

EGC-2

concentration to 1.0 mM.

11. Turn on the

EGC-1

and

EGC-2

currents.

12. Run at the selected settings until the mixer is filled with the eluent

mixture. At a flow rate of 1 mL/min it will take about 45 minutes to
fill the 4-mm mixer and about 15 minutes to fill the 2-mm mixer.

NOTE The void volume of the 4-mm mixer is about

16 mL. The void volume of the 2-mm mixer is
about 5 mL.

13. Turn off the

EGC-1

and

EGC-2

currents.

14. Turn off the pump.

15. Remove the backpressure coil from the

TO INJ VALVE IN - P/ELUENT

OUT

line and the mixer inlet. Reconnect the two lines.

16. Disconnect the mixer outlet from the

TO SRS/AES REGEN OUT

line

and remove the 10-32 to 1/4-28 coupler.

To avoid damaging the suppressor and auxiliary electrolytic device (if
installed), always turn them off before filling the mixer. The pump flow
is on during filling; however, no flow reaches the suppressor or
auxiliary electrolytic device.