ICS-2100 Ion Chromatography System

250

Doc. 065291-01 3/09

Timed Event Notes:

At time INIT, the injection valve is set to Load.

At time 0.00, the injection valve is set to Inject, an autozero is performed
(TTL2=1), the data acquisition mark is sent (RLY1=1), and the pump is
turned on (RLY2=1).

At time 0.5, TTL2 and RLY1 are switched back to 0 to be ready for the next
run. RLY2 remains at 1 to keep the pump on.

TTL1 does not need to be set in the timed events because the Load and Inject
commands control the ICS-2100 injection valve automatically (see

“Enable

AS Control of the Injection Valve” on page 248

).

To run the above method, set up an AS schedule. The example in

Figure C-11

includes a four-point calibration (Lines 1–4) followed by two samples (Lines 5
and 6).

Figure C-11. Example AS Schedule for Stand-Alone Operation

Help Prompt

SAVE TO

RUN

SCHEDULE

25.0

5.0

1

1

1

4

3

5

5

1

6

6

1

1

MISSING VIAL ACTION:

INJ/

INJ

VIAL

METHOD

VOL(uL)

VIAL#

LINE

END

START

EDIT

SKIP

1

1

1

1-4

5

6

-

-

-

-

25.0