ICS-2100 Ion Chromatography System

148

Doc. 065291-01 3/09

17. Slide the EluGen cartridge down

into the holder until it stops.
Make sure the cable and
backpressure coil remain out of
the holder (see

Figure 5-42

).

NOTE To avoid crimping the

lines inside the holder, do
not twist the cartridge
before you slide it into
the holder. The lines are
designed to coil inside the
holder.

Figure 5-42. Installing the EluGen Cartridge in the Holder

18. Orient the cartridge’s

blue 4-pin electrical
cable connector as
shown in

Figure 5-43

.

Push it onto the

EGC-1

connector on the ICS-
2100. Twist the ring on
the cable connector
fingertight to secure it.

Figure 5-43. Connecting the EluGen Cartridge Cable