The outputs can be used to control functions in external devices such as an
autosampler or another Dionex module. When connected to a controlling device,
the inputs can be programmed to perform the following ICS-2100 functions:

Switch the injection valve position (load/inject)

Perform an autozero command (set the conductivity to zero)

Turn the pump on and off (also turns the suppressor on and off)

Automatically turn on the optional auxiliary power supply simultaneously
with the pump flow

Send a chart mark signal to the analog output. The mark is 10% of the full-
scale voltage, and the duration is 0.5 seconds. A mark can be used, for
example, to indicate the injection.

Relay outputs 1 and 2 can be programmed to switch any low-voltage control.
Switched current must be less than 200 mA and 42 V peak.

Connecting a TTL or Relay

1. Locate the twisted pair of

wires (P/N 043598) and the
12-position connector plug
(P/N 923687) (see

Figure C-2

) in the ICS-2100

Ship Kit (P/N 064375).

2. Follow these basic steps to

connect the TTL or relays.

a.

For each relay or TTL to
be used, connect an
active wire (red) and a ground wire (black) to the 12-position connector
plug at the appropriate pin locations. Refer to

Figure C-1

or the label on

the ICS-2100 rear panel for the connector pin assignments.

To attach a wire to the plug, strip the end of the wire, insert it into the
plug, and use a screwdriver to tighten the locking screw. If necessary,
multiple ground wires can be attached to a single TTL input/output
ground pin.

When attaching wires to the connector plug, be careful not to allow
stray strands of wire to short to the adjoining position on the
connector.

Figure C-2. 12-Position Connector Plug

Position 1

Position 12

Locking
Screws