This error occurs if you turn on the suppressor and the ICS-2100 cannot
establish a connection with the suppressor.

To troubleshoot:

1. Check the suppressor cable connection (see

Section 5.11

).

2. If the error persists, contact Dionex for assistance. There may be a

problem in the suppressor controller electronics. The ICS-2100
electronics components cannot be serviced by the user.

This error may be caused by an expended or dirty suppressor or by a
malfunction in the suppressor controller electronics.

To troubleshoot:

1. Follow the instructions in the suppressor manual to regenerate the

suppressor. Suppressor manuals are included on the Dionex Reference
Library CD-ROM (P/N 053891).

2. Follow the instructions in the suppressor manual to clean the suppressor.

3. If a malfunction in the suppressor controller is suspected, contact Dionex

for assistance. The ICS-2100 electronics components cannot be serviced
by the user.

SUPPRESSOR NOT CONNECTED

SUPPRESSOR OVER CURRENT