Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 178

Advertising
background image

ICS-2100 Ion Chromatography System

166

Doc. 065291-01 3/09

10. Check that none of the lines are

caught under the CR-TC and
then push the CR-TC onto the
metal stud inside the EGC
holder. Push down until the
CR-TC snaps into place (see

Figure 5-60

).

Figure 5-60. CR-TC Installed in Holder

11. Feed the CR-TC cable through

the slot on the side of the EGC
holder (see

Figure 5-61

).

12. Turn the EGC holder right-side

up.

Figure 5-61. CR-TC Cable

13. Reconnect the

EGC IN

and

EGC OUT

lines to the EluGen cartridge inlet and

outlet ports.

14. Turn over the EluGen cartridge (fittings facing down). Check for bubbles in

the eluent generation chamber; if necessary, shake and tap the EluGen
cartridge to remove bubbles.

CR-TC

Cable