5 • Service

Doc. 065291-01 3/09

163

1. Locate the yellow 0.5 mL/min, 7 MPa (1000 psi) backpressure coil

(P/N 053765) and 10-32 to 1/4-28 coupler (P/N 042806) in the ICS-
2100 Ship Kit (P/N 064375).

2. Disconnect the

TO INJ VALVE IN - P/ELUENT OUT

line from the EGC

CO

3

Mixer inlet.

3. Connect one end of the backpressure coil to the

TO INJ VALVE IN -

P/ELUENT OUT

line.

4. Connect the 10-32 to 1/4-28 coupler to the free end of the

backpressure coil.

5. Disconnect the

TO SRS/AES REGEN OUT

line from the suppressor

REGEN OUT

port.

6. Connect the

TO SRS/AES REGEN OUT

line to the 10-32 to 1/4-28

coupler (see

Figure 5-58

).

NOTE The backpressure coil connection in

Figure 5-58

is temporary; it is required only during the
conditioning procedure described in the next
section.

Figure 5-58. Backpressure Coil Connections for Conditioning the EPM

with Carbonate/Bicarbonate Eluent

10-32 to

1/4-28

coupler

TO INJ
VALVE
IN - P/
ELUENT OUT

Backpressure
Coil

TO SRS/AES

REGEN OUT

10-32 to
10-32 coupler