5 • Service

Doc. 065291-01 3/09

149

19. Remove the plug from the EluGen cartridge vent opening and

connect the EluGen cartridge

VENT

line.

Recording the EluGen Cartridge Serial Number in Chromeleon

The serial number is printed on the EluGen cartridge label.

1. Open the Chromeleon Server Configuration program (select Start >

All Programs > Chromeleon > Server Configuration).

2. Double-click the ICS-2100 system icon. The properties dialog box for

the ICS-2100 appears.

3. Click the Eluent Generator tab (see

Figure 5-44

).

4. Select the EGC-1 check box.

5. Enter the EluGen cartridge serial number and click OK.

Figure 5-44. Eluent Generator Configuration Properties Example