Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 180

Advertising
background image

ICS-2100 Ion Chromatography System

168

Doc. 065291-01 3/09

1. Locate the 10-32 to 1/4-28 coupler (P/N 042806) in the ICS-2100

Ship Kit (P/N 064375).

2. Use the coupler to connect the line labeled

TO INJ VALVE IN -

P/ELUENT OUT

and the line labeled

SRS/AES REGEN OUT

.

3. Verify that the current to the suppressor and the auxiliary electrolytic

device (if installed) is turned off.

4. Verify that the

Automatically turn on with pump

option is not

selected on the ICS-2100 touch screen

ELUENT GENERATOR

page

(see

Section B.8

). Disabling this option lets you turn on the pump,

without automatically turning on the CR-TC.

5. Turn on the pump and set the flow to the rate recommended for your

application. Pump deionized water through the CR-TC for at least
10 minutes.

6. Turn off the pump and disconnect the coupler.

7. Reconnect the line labeled

TO INJ VALVE IN - P/ELUENT OUT

to

port (

P

)

2

on the injection valve.

8. Reconnect the line labeled

TO SRS/AES REGEN OUT

to the

REGEN

OUT

port on the suppressor.

9. Reconnect the CR-TC and EluGen cartridge cables.

10. If you disabled the

Automatically turn on with pump

option in

Step 3

, select the option.

To avoid damaging the suppressor and auxiliary electrolytic device (if
installed), always turn them off before hydrating the CR-TC. The pump
flow is on during hydrating; however, no flow reaches the suppressor
or auxiliary electrolytic device.