10. Remove the new EluGen cartridge from the shipping box.

11. Orient the EluGen

cartridge with the fittings
facing up and remove the
plugs from the EluGen
cartridge

INLET

and

OUTLET

ports (see

Figure 5-50

).

Figure 5-50. EluGen Cartridge Inlet and Outlet Ports

12. Set the EluGen cartridge in the service area on top of the ICS-2100

and orient it with the cable toward the EGC holder (see

Figure 5-51

).

13. Connect

EGC IN

to the EluGen cartridge

INLET

port.

14. Connect

EGC OUT

to the EluGen cartridge

OUTLET

port.

Figure 5-51. EluGen Cartridge in Service Area

OUTLET

INLET

EluGen

Cartridge

Cable