ICS-2100 Ion Chromatography System

80

Doc. 065291-01 3/09

This error indicates a problem in the ICS-2100 electronics.

To troubleshoot:

Contact Dionex for assistance. The ICS-2100 electronics components cannot
be serviced by the user.

The leak sensor is installed in the drip tray at the bottom of the component
panel (see

Figure 2-8

). If liquid accumulates in the tray, the sensor signals the

problem and this error message appears.

To troubleshoot:

1. Locate the source of the leak by visually inspecting the tubing, fittings,

and components on the component panel. Refer to

Section 4.3

.

2. Tighten fittings or replace tubing and fittings as required. Refer to

Section 4.3

for detailed troubleshooting of various types of leaks.

3. After fixing the leak, dry the drip tray thoroughly to prevent the leak

sensor from triggering additional error messages.

If the injection valve fails to switch position within 1 second of being toggled,
the ICS-2100 Moduleware reports an error to Chromeleon and this error
message appears.

To troubleshoot:

1. If a sequence is being executed, terminate the sequence by selecting

Batch > Stop in Chromeleon.

2. Try to toggle the valve from Load to Inject by clicking the Load

and

Inject buttons on the ICS-2100 Control panel in Chromeleon or on the
touch screen

HOME

page.

HARDWARE NOT PRESENT

LEAK SENSOR WET

LOAD/INJECT VALVE ERROR