5 • Service

Doc. 065291-01 3/09

167

15. Reinstall the EluGen cartridge in the holder.

16. Remove the cap on the luer adapter by twisting. Save the cap.

17. Twist the luer lock and

EGC VENT

line into the luer adapter.

18. Do not connect the EluGen cartridge cable until after hydrating the

CR-TC.

Hydrating the CR-TC

The CR-TC must be hydrated before initial use by running ASTM
filtered, Type 1 (18 megohm-cm) deionized water through the CR-TC for
10 minutes, while bypassing the columns and suppressor.

Figure 5-62

illustrates the liquid flow path through the CR-TC during the

hydration procedure.

Figure 5-62. CR-TC Plumbing for Hydration