Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 31

Advertising
background image

2 • Features

Doc. 065291-01 3/09

19

Pressure Transducer

The pressure transducer measures the system backpressure.

Leak Sensor

The leak sensor is installed in the drip tray at the bottom of the component
panel. If liquid accumulates in the tray, an error message is logged in the
Chromeleon Audit Trail and displayed on the LCD touch screen.

Pump Heads

The ICS-2100 includes a dual-piston serial pump. The flow rate can be set
to 0.00 mL/min or to between 0.05 and 5.00 mL/min. However, for
optimum performance, set the flow rate to between 0.40 and
2.00 mL/min. Setting the flow rate to 0.00 mL/min turns off the pump.
See

Section 2.4.3

for details about the pump.

Injection Valve

The injection valve is a six-port, electrically-activated Rheodyne valve. A
25-

μL sample loop (P/N 042857) is installed on the valve at the factory.

See

Section 2.4.6

for details about valve operation.

Auxiliary Valve (Optional)

The auxiliary valve is a two-position, electrically-activated Rheodyne
valve (6-Port Valve Kit, P/N 069472; 10-Port Valve Kit, P/N 069473).
See

Section 2.4.6

for details about valve operation.

DS6 Heated Conductivity Cell

The flow-through conductivity cell measures the electrical conductance
of analyte ions as they pass through the cell. A heat exchanger inside the
cell regulates the temperature, which can be set to between 30 and 55 °C.
For optimum performance, set the temperature to at least 7 °C above the
ambient temperature and 5 °C above the column oven temperature. See

Section 2.4.10

for details about the cell.