Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 255

Advertising
background image

C • TTL and Relay Control

Doc. 065291-01 3/09

243

To select TTL input functions and the control type from the touch screen,
go to the

INPUT/OUTPUT

touch screen page (see

Figure C-4

).

TTL Input Control Types

The ICS-2100 TTL inputs respond to four types of
signals to accommodate different controlling devices.
The default control type,

Normal Edge

, is compatible

with the output signals provided by Dionex modules.

If the device connected to the ICS-2100 does not send a normal edge
signal, select the appropriate control type. Refer to the documentation
provided with the controlling device and the information below to select
the correct type.

Normal Edge: In normal edge
operation, the negative (falling)
edge of a signal turns on the
function. For example, for the

Load/Inject

function, the

negative edge switches the
injection valve position to Load.

Figure C-4. Input/Output Page: Selecting TTL Functions