4. Remove the EluGen

cartridge

VENT

line

(see

Figure 5-35

)

from the EluGen
cartridge by
unscrewing the luer
lock from the luer
adapter.

If necessary, use a
wrench to hold the
luer adapter in place
while unscrewing
the vent line.

Figure 5-35. EluGen Cartridge Vent Line and Luer Fitting

5. Install the luer cap (P/N 053981), provided with the EluGen cartridge,

onto the luer adapter. This prevents leaks from the vent opening when
you turn over the EluGen cartridge

(see Step 7)

.

6. Lift the EluGen cartridge straight up and out of the cartridge holder.

7. Turn the cartridge upside down and set it in the EluGen cartridge

service area on top of the ICS-2100 (see

Figure 5-36

and

Figure 5-37

).

Figure 5-36. EluGen Cartridge Service Area

Luer Lock

Luer Adapter

EGC
Vent
Line

EluGen Cartridge

Area