Conditioning the EluGen Cartridge

Always condition the EluGen cartridge before initial use. This requires
directing the backpressure coil connected to the EluGen cartridge outlet to
waste, and then generating 50 mM of eluent at 1.00 mL/min for
30 minutes. Follow the steps below:

1. Set a temporary waste container, such as a beaker, on top of the ICS-

2100 and direct the yellow backpressure coil from the EluGen
cartridge

OUTLET

port to the container.

2. From the Chromeleon Control panel or the ICS-2100 touch screen

HOME

page, set the pump flow rate to 1.00 mL/min and set the

EluGen cartridge concentration to 50 mM.

3. Verify that the suppressor, the auxiliary electrolytic device attached to

the

EGC-2

connector (if installed), and the CR-TC are off.

4. Verify that the

Automatically turn on with pump

options are not

selected on the touch screen

SUPPRESSOR

page (see

Section B.9

)

and

ELUENT GENERATOR

page (see

Section B.8

). Disabling these

options lets you turn on the pump, without automatically turning on
the suppressor and CR-TC.

5. Verify that the eluent reservoir is filled with ASTM filtered, Type 1

(18 megohm-cm) deionized water.

6. Turn on the pump.

7. Run at the selected settings (1.00 mL/min at 50 mM) for 30 minutes.

8. Turn off the pump.

9. Disconnect the EluGen cartridge cable.

10. Remove the backpressure tubing from the waste container and

remove the waste container.

To avoid damaging the suppressor, auxiliary electrolytic device (if
installed), and CR-TC, always turn them off before conditioning the
EluGen cartridge. The pump flow is on during conditioning; however,
no flow reaches the suppressor, auxiliary electrolytic device, or CR-
TC.