5 • Service

Doc. 065291-01 3/09

189

5. Check that none of the

lines are caught under the
CR-TC and then push the
CR-TC onto the metal
stud inside the EGC
holder. Push down until
the CR-TC snaps in place
(see

Figure 5-78

).

Figure 5-78. CR-TC Installed in Holder

6. Feed the CR-TC cable

through the slot on the
side of the EGC holder
(see

Figure 5-79

).

Figure 5-79. CR-TC Cable

7. Turn the EGC holder right side up and set it on top of the ICS-2100.

8. Direct the

WASTE, GAS SEPARATOR

line from the top of the ICS-

2100 to the rear panel. Snap the line onto one of the tubing clips on
the rear panel. Connect the line to the gas separator waste tube (see

Section B.10.1

).