1. Disconnect the

TO INJ VALVE IN - P/ELUENT OUT

line from the EGC

CO

3

Mixer inlet.

2. Locate the yellow 0.5 mL/min, 7 MPa (1000 psi) backpressure coil

(P/N 053765) and a 10-32 to 10-32 coupler (P/N 042627) in the ICS-
2100 Ship Kit (P/N 064375).

3. Use the coupler to connect the

TO INJ VALVE IN - P/ELUENT OUT

line

to one end of the backpressure coil (see

Figure 5-56

).

4. Direct the free end of the backpressure coil to a temporary waste

container.

Figure 5-56. Plumbing Connections for Conditioning the

EluGen Cartridge with Carbonate Eluent

EGC Cable

Temporary

Waste

Container

Back-

pressure

Coil

TO INJ VALVE

IN - P/

ELUENT OUT