5 • Service

Doc. 065291-01 3/09

183

1. Locate the following items in the ICS-2100 Ship Kit (P/N 064375): a

yellow 0.5 mL/min, 7 MPa (1000 psi) backpressure coil
(P/N 053765), a 10-32 to 10-32 coupler (P/N 042627), and a 10-32 to
1/4-28 coupler (P/N 042806).

2. Use the 10-32 to 10-32 coupler to connect one end of the

backpressure coil to the

TO INJ VALVE IN - P/ELUENT OUT

line (see

Figure 5-72

).

3. Connect the other end of the backpressure coil to the coupler on the

mixer inlet.

4. Use the 10-32 to 1/4-28 coupler to connect the mixer outlet to the

TO

SRS/AES REGEN OUT

line.

NOTE The backpressure coil connection is temporary;

it is used only during initial filling of the mixer.

Figure 5-72. Mixer Connected Directly to Eluent

Generator Outlet for Initial Filling

Back-

pressure

Coil

TO INJ PORT IN - P/ELUENT OUT

Mixer Inlet

Mixer Outlet

TO SRS/AES
REGEN OUT