Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 176

Advertising
background image

ICS-2100 Ion Chromatography System

164

Doc. 065291-01 3/09

Conditioning the EluGen Cartridge

Always condition the EluGen cartridge for 30 minutes before initial use.

1. Make sure the eluent reservoir is filled with ASTM filtered, Type I

(18-megohm) deionized water.

2. Verify that the suppressor and the auxiliary electrolytic device (if

installed) are off.

3. Verify that the

Automatically turn on with pump

option is not

selected on the ICS-2100 touch screen

SUPPRESSOR

page (see

Section B.9

). Disabling this option lets you turn on the pump, without

automatically turning on the suppressor.

4. From the ICS-2100 Control panel in Chromeleon or the touch screen

HOME

page, set the pump flow rate to 1.00 mL/min.

5. Turn on the pump.

6. Select the following settings from the Chromeleon Control panel or

the touch screen

ELUENT GENERATOR

page:

Set the

EGC-1

concentration to 9 mM.

Set the

EGC-2

concentration to 1 mM.

7. Turn on the EGC-1 and EGC-2 currents.

8. Run at these settings for 30 minutes.

9. Turn off the EGC-1 and EGC-2 currents.

10. Turn off the pump.

11. Remove the backpressure coil.

12. Reconnect the

TO INJ VALVE IN - P/ELUENT OUT

line to the mixer

inlet.

13. Reconnect the

TO SRS/AES REGEN OUT

line to the suppressor

REGEN

OUT

port.

To avoid damaging the suppressor (and auxiliary electrolytic device, if
installed), always turn them off before conditioning the EluGen
cartridge. The pump flow is on during conditioning; however, no flow
reaches the suppressor or auxiliary electrolytic device.