Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 87

Advertising
background image

4 • Troubleshooting

Doc. 065291-01 3/09

75

This error occurs when the device connected to the auxiliary power supply is
attempting to operate at a higher voltage than is allowed (35 V).

To troubleshoot:

1. Check the cable connecting the device to the power supply for damage or

a faulty connection.

2. If the error persists, refer to the manual provided by the device vendor for

assistance.

When the pump flow stops, the auxiliary power supply automatically turns off
(to prevent damage to the electrolytic device) and this error occurs.

To troubleshoot:

If the pump stopped unexpectedly, see

Section 4.6

for pump troubleshooting

information.

This error occurs when the column heater temperature is higher than the
maximum allowed. This error may occur if the ICS-2100 is operating in an
extreme environment (greater than 40

°

C (104

°

F)).

To troubleshoot:

Refer to

Section A.3

for environmental specifications.

AUXILIARY POWER SUPPLY OVER VOLTAGE

AUXILIARY POWER SUPPLY STOPPED FOR ZERO FLOW
RATE ERROR

COLUMN HEATER EXCEEDS SAFE TEMPERATURE