Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 260

Advertising
background image

ICS-2100 Ion Chromatography System

248

Doc. 065291-01 3/09

Assign ICS-2100 Input Functions

Go to the ICS-2100

INPUT/OUTPUT

touch screen page and verify that the

Inject

function is assigned to TTL input 1, that Autozero is assigned to TTL input 2, and
that the

TTL Input Control Type

is

Normal Edge

(see

Figure C-7

).

Enable AS Control of the Injection Valve

1. On the AS front panel, press

Menu

and

0

to go to the

TIME FUNCTION OUT

screen. In the

LOAD/INJECT VALVE: TTL 1

field, select

ENABLED

.

2. Turn the AS power off and then back on.

You can now issue commands from the AS to switch the ICS-2100 injection
valve position.

Figure C-7. Input/Output Page

Figure C-8. AS Time Function Out Screen

Help Prompt

TIME FUNCTION OUT

TTL

1

2

LOAD/INJECT VALVE:

LOAD/INJECT VALVE:

ENABLED

DISABLED

STATE