Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 189

Advertising
background image

5 • Service

Doc. 065291-01 3/09

177

4. Connect the pump outlet to the mixer inlet (see

Figure 5-69

).

5. Disconnect the mixer outlet line (labeled

TO INJ VALVE PORT P

) from

port

P

(

2

) on the injection valve.

6. Direct the mixer outlet line to a temporary waste container, such as a

beaker.

7. From the ICS-2100 Control panel in Chromeleon or the ICS-2100

touch screen

HOME

page, set the pump flow rate to 5.00 mL/min.

8. Verify that the suppressor and the auxiliary electrolytic device (if

installed) are off.

9. Verify that the

Automatically turn on with pump

option is not

selected on the touch screen

SUPPRESSOR

page (see

Section B.9

).

Disabling this option lets you turn on the pump, without
automatically turning on the suppressor.

10. Turn on the pump.

Figure 5-69. Pump Outlet Connected to Mixer Inlet for Mixer Hydration

To avoid damaging the suppressor and auxiliary electrolytic device (if
installed), always turn them off before filling the mixer. The pump flow
is on during filling; however, no flow reaches the suppressor or
auxiliary electrolytic device.

Mixer Inlet

Pump Outlet